Installing and Controlling Third-Party Devices
Soma 3710 UV Detector: Overview

 

Device Type:

UV detector

Device Driver:

Soma: Soma_UV-Vis_Detector_S-3710_178
(up to firmware version 1.78)
Soma: Soma_UV-Vis_Detector_S-3710_New

(from firmware version 1.79 on)

 

What is required?

License:

Timebase Class 2

Connection:

Null modem cable (25-pins on the instrument's side, for more information, refer to  Null Modem Cable in the Appendix of the Hardware Installation section.)

Hardware
Prerequisites:

  • An A/D converter is required for data acquisition. A UCI Universal Chromatography Interface including the appropriate signal cables may be used.

  • A free RS-232 port is required for control.

Control Panel:

UV-Fluorescence.pan

 

Further Information

For information about how to install the detector, refer to  Soma 3710 UV Detector: Installation.

For an overview of manufacturers whose instruments can be controlled by Chromeleon, refer to  Installing and Controlling Third-Party Devices.